Page 1 - Maintenance Manual

ADS Intrinsically-Safe FlowShark Installation, Operation, and Maintenance Manual March 2011 QR 775003 A2 4940 Res

Page 2

1-4 ADS FlowShark IS Manual Note: A legacy Rechargeable Battery Pack (RBP) may be connected to the COMM connector on the FlowShark IS in a similar

Page 3 - Table of Contents

Sensor Installation and Connection 3-43 Securing the Sensor Cables in the Pipe and Manhole Securing the sensor cables from the ring (or band, when

Page 4

3-44 ADS FlowShark IS Manual 2. Secure the cables with a cable tie and plastic anchor every 18 to 24 inches along the pipe crown (when necessary) f

Page 5 - Chapter 8 Maintenance and

Sensor Installation and Connection 3-45 Connecting the Sensors to the Monitor After installing the ring and sensors and securing the sensor cables

Page 6

3-46 ADS FlowShark IS Manual 5. Connect each sensor cable to the appropriate port. Tighten each connector in a clockwise direction until it clicks

Page 7 - Introduction

4-1 C H A P T E R 4 Communication After installing the sensors in the pipe and connecting the sensors to the ADS® FlowShark® IS monitor, it is nece

Page 8 - Intrinsic Safety

4-2 ADS FlowShark IS Manual Installing the Monitor Communication Hardware Establishing remote, wireless communications with the FlowShark IS monitor

Page 9

Communication 4-3  Verify that the communication cable trench will not damage an existing sprinkler system. Therefore, locate any sprinkler lines.

Page 10 - Additional Features

4-4 ADS FlowShark IS Manual Gathering Tools and Materials Obtain the following items to perform the installation:  Ratchet set with 7/16-inch socke

Page 11

Communication 4-5 Wire ends stripped ¼-inch  Cut off the old stripped ends (when applicable) back to the jacket.  Use the small side cutters to

Page 12

4-6 ADS FlowShark IS Manual Forming a thin taper using vinyl tape 4. Excavate a trench at least 12 inches (31 cm) deep and 4 inches (10 cm) wide fr

Page 13 - Maintenance Restrictions

Introduction 1-5  Only limited field repairs to the FlowShark IS are allowed; all other repairs must be performed by an ADS authorized technician

Page 14 - Compliance, and Conformity

Communication 4-7  Use urethane foam to seal the space between the conduit and the corbel or manhole wall to prevent infiltration into the manhole.

Page 15 - FCC Compliance

4-8 ADS FlowShark IS Manual Installing the Wireless EMU Installing the wireless EMU primarily involves wiring the communication cable to the Communic

Page 16 - GSM/GPRS Modem Information

Communication 4-9 Gathering Tools and Materials Obtain the following items to perform the installation:  Medium (No. 2) Phillips head screwdriver 

Page 17 - R&TTE-CE Information

4-10 ADS FlowShark IS Manual Inside component view of EMU box 1. Unscrew the 4 large plastic screws on the cover of the EMU box by hand and remove

Page 18

Communication 4-11 3. Loosen the nut on the communication cable entry fitting. If the fitting is snapped closed, use a small flat-blade screwdriver

Page 19 - Sira certification label

4-12 ADS FlowShark IS Manual 5. Terminate the 12 wires into the orange connector in the following way using the ⅛-inch flat screwdriver. This inform

Page 20

Communication 4-13 Connecting the communication cable to port on the Communication Interface/Barrier Box 7. Tighten the nut on the communication ca

Page 21 - Introduction 1-15

4-14 ADS FlowShark IS Manual SIM card seated in EMU modem 10. (applies only to battery-powered EMUs) Re-install the battery pack in the EMU in th

Page 22

Communication 4-15 Installing the EMU Installing the EMU involves mounting the EMU inside the customer-supplied enclosure, connecting the communicati

Page 23 - Introduction 1-17

4-16 ADS FlowShark IS Manual 4. (applies only when using external power) Run and connect the external power cable to the power supply, and then con

Page 24

1-6 ADS FlowShark IS Manual • Only ADS approved sensors (ADS p/n 106186, 106194, and 3754-0001/0002/0003) shall be connected to the ADS Model 5000-

Page 25 - Introduction 1-19

Communication 4-17 Connecting to the Monitor in the Field Communicating with the monitor on-site occurs by connecting directly to the monitor from a

Page 26

4-18 ADS FlowShark IS Manual Direct connection using the IS PC communication cable 3. Use the Profile software to establish communication with the

Page 27 - Introduction 1-21

Communication 4-19 EMU serial cable (ADS p/n 3800-0147) Communicate with the monitor through the EMU in the following way: 1. Disconnect the ribbo

Page 28

4-20 ADS FlowShark IS Manual 3. Connect the other (metal) end of the cable to the serial port on the laptop computer. Direct connection using EMU s

Page 29 - Introduction 1-23

5-1 C H A P T E R 5 External Power The ADS® FlowShark® IS system (monitor and EMU) can receive power from an external DC power source connected to

Page 30

5-2 ADS FlowShark IS Manual Typical Installation A typical external power installation includes the following:  A single-phase, 120-volt AC power l

Page 31 - Introduction 1-25

External Power 5-3 Installing an ADS Power Supply The ADS 12VDC Power Supply (p/n 5000-0727) is a universal input supply that automatically adjusts t

Page 32

5-4 ADS FlowShark IS Manual Note: Do not drill new holes through the housing for mounting! This will expose the internal components of the enclosure

Page 33 - Introduction 1-27

External Power 5-5 Wiring the external power cable to the ADS Power Supply 6. Verify the following before connecting to the power source:  Batter

Page 34

5-6 ADS FlowShark IS Manual 9. Securely tighten the cable grip fittings and replace the enclosure cover. Make sure the cover properly aligns with th

Page 35 - Introduction 1-29

Introduction 1-7 Maintenance Restrictions As mentioned earlier, all ADS FlowShark IS flow monitors are manufactured to meet IS standards. ADS can v

Page 36

6-1 C H A P T E R 6 Configuration and Activation After installing the ADS® FlowShark® IS flow monitor and sensors and setting up communication, i

Page 37 - Introduction 1-31

6-2 ADS FlowShark IS Manual Configuring the Monitor Location To ensure the most efficient and useful results for each monitor location, the user mus

Page 38 - Declaration of Conformity

Configuration and Activation 6-3 Once installed, start the Profile software from an office or field computer by selecting Start > Programs &g

Page 39 - Introduction 1-33

6-4 ADS FlowShark IS Manual Creating a Monitor Location Creating a new location includes entering and setting up the location information in the fol

Page 40

Configuration and Activation 6-5 3. Enter the new Location Name. Monitor names must follow this convention depending on length:  Use unique n

Page 41 - Product Warranty

6-6 ADS FlowShark IS Manual Note: The start and end times displayed in the Monitor Data section represent the complete date/time range of flow data

Page 42 - Shipping

Configuration and Activation 6-7 1. Select the monitoring point for which you want to create the installation. Monitoring Point 1 selected fro

Page 43 - System Overview

6-8 ADS FlowShark IS Manual Create Installation – Installation dialog 4. Select the Pipe radio button, and then click on the Next button. Create

Page 44

Configuration and Activation 6-9 Create Installation – Dimension/Parameters dialog 6. Enter the proper pipe dimensions in the corresponding fi

Page 45

6-10 ADS FlowShark IS Manual 7. Enter an appropriate name for the installation in the text field, and then click on the Next button. Create Instal

Page 46 -  Voltage Regulator Board

1-8 ADS FlowShark IS Manual Warnings, Certifications, FCC Compliance, and Conformity Manhole and sewer system work involves confined space entry and

Page 47 - Digital Signal Processor

Configuration and Activation 6-11 Installation Generator dialog 10. Select File > Exit to close the Installation Generator. Selecting and E

Page 48 - DSP Board

6-12 ADS FlowShark IS Manual Devices selected for HSV4 3. Select the Edit > Properties option or click on the Properties toolbar button. Prope

Page 49 - Analog Board

Configuration and Activation 6-13 device in the Monitoring Point Devices section to ensure Profile includes the device in the LIF. Deselect the

Page 50 - Voltage to Frequency Board

6-14 ADS FlowShark IS Manual Editing the Ultrasonic Device Edit the ultrasonic device in the following way, and then click on the OK button when com

Page 51 - Voltage Regulator Board

Configuration and Activation 6-15 from the sensor face to the bottom of the pipe. This (default) setting effectively represents the vast majori

Page 52 - Connectors

6-16 ADS FlowShark IS Manual Note: Typically, the velocity device should not require editing. However, when editing is necessary, ADS recommends ac

Page 53

Configuration and Activation 6-17  Max Velocity Select the appropriate option from this drop-down list to fine tune the maximum velocity valu

Page 54 - Sensors

6-18 ADS FlowShark IS Manual Editing the Pressure Device Edit the pressure device in the following way, and then click on the OK button when complet

Page 55 - Ultrasonic Depth Processing

Configuration and Activation 6-19  Access the ADS website at www.adsenv.com on the Internet.  Select ADS Environmental Services > Support

Page 56 - IS Pressure Depth Sensor

6-20 ADS FlowShark IS Manual Setting the Communication Parameters Setting the communication parameters involves designating the communication ports,

Page 57 - IS Doppler Velocity Sensor

Introduction 1-9 Control Drawing This drawing depicts the interconnections allowed for the FlowShark IS. It is intended for use by inspection profe

Page 58 - Connection

Configuration and Activation 6-21 2. Select Communication Parameters from the Functions drop-down list, and click on the Perform button. The Se

Page 59

6-22 ADS FlowShark IS Manual  Attempts Enter the number of times you want your local PC to request data from the monitor following a failed attem

Page 60 - Flow Hydraulics

Configuration and Activation 6-23 an external power source, below which you want Profile to provide notification for the FlowShark IS monitor. A

Page 61

6-24 ADS FlowShark IS Manual Activating the Monitor After configuring the monitor, activate the monitor to initiate the flow monitoring process base

Page 62

Configuration and Activation 6-25 Monitor Diagnostics dialog 2. Click on the Connect button to establish communication with the monitor. Profi

Page 63 - Standard Installation

6-26 ADS FlowShark IS Manual Running Device Diagnostics Profile’s diagnostics tool enables the user to verify the proper operation of sensors, obtai

Page 64

Configuration and Activation 6-27 2. Click on the Connect button to establish communication with the monitor. Profile initiates communication w

Page 65 - Assembling the Ring

6-28 ADS FlowShark IS Manual Diagnostics dialog showing results from ultrasonic depth sensor firing  Click on the Advanced button to view more d

Page 66

Configuration and Activation 6-29  Click on the Close button to exit the Ultrasonic 1 device dialog and return to the Monitor Diagnostics dial

Page 67

6-30 ADS FlowShark IS Manual Diagnostics dialog showing results from velocity sensor firing  Click on the Advanced button to view more detailed d

Page 68

1-10 ADS FlowShark IS Manual These limits are designed to provide reasonable protection against harmful interference in a residential installation.

Page 69

Configuration and Activation 6-31  Click on the Close button to exit the Velocity 1 device dialog and return to the Monitor Diagnostics dialog

Page 70

6-32 ADS FlowShark IS Manual Diagnostics dialog showing results from pressure depth sensor firing  Click on the Advanced button to view more deta

Page 71 - Attaching the spreader bars

Configuration and Activation 6-33 7. Click on the Disconnect button to discontinue communication with the monitor when finished running diagnos

Page 72

6-34 ADS FlowShark IS Manual Collecting Data from the Monitor The data collection function available through Diagnostics provides enhanced features

Page 73

Configuration and Activation 6-35 Monitor Diagnostics dialog 2. Click on the Connect button to establish communication with the monitor. Profi

Page 74

6-36 ADS FlowShark IS Manual Collect the Specified Data dialog 4. Designate the range of data you want to collect from the monitor by editing the

Page 75

Configuration and Activation 6-37 Collect The Specified Data dialog indicating data collection was a success 6. Click on the Next button. The

Page 76

6-38 ADS FlowShark IS Manual View Hydrograph – [location name] dialog 7. Review the data on the hydrograph, and then click on the Next button. Und

Page 77

Configuration and Activation 6-39 View Depth and Velocity Scattergraph – [location name] dialog 8. Review the data on the scattergraph, and th

Page 78 - Sensor cabling

6-40 ADS FlowShark IS Manual View Ultrasonic and Pressure Scattergraph – [location name] dialog 9. Review the data on the scattergraph, and then c

Page 79

Introduction 1-11 R&TTE-CE Information U.S. Hazardous Area Compliance The FlowShark IS has been certified for use in Class I, Division 1, Gr

Page 80

Configuration and Activation 6-41 Site DR Analysis – [location name] dialog 10. Review any issues identified and the suggested actions, and th

Page 81

6-42 ADS FlowShark IS Manual Upgrading the Monitor Firmware Profile enables you to download updated firmware to the FlowShark IS flow monitor that m

Page 82 - Special Installations

Configuration and Activation 6-43 Monitor Diagnostics – [location name] dialog 2. Click on the Connect button to establish communication with

Page 83 - Gathering Parts and Supplies

6-44 ADS FlowShark IS Manual Form Firmware Download dialog 4. Click on the Browse button corresponding to DSP file to locate and designate the DSP

Page 84

Configuration and Activation 6-45 Viewing Diagnostic and Data Logs Profile generates detailed logs for many activities performed through Diagnos

Page 85

6-46 ADS FlowShark IS Manual 2. Select the monitoring point for which you want to view a log from the Monitoring Point drop-down list. 3. Select L

Page 86 - Standard Ultrasonic Mount

Configuration and Activation 6-47 Example of Monitor Status log 7. (optional) Click on the Print button to print the contents of the log file

Page 87

7-1 C H A P T E R 7 Monitor Installation After installing the sensors and establishing communications, install the ADS® FlowShark® IS monitor in t

Page 88

7-2 ADS FlowShark IS Manual Mounting the Monitor to the Manhole Wall Mount the monitor to the manhole wall using the monitor bracket in the following

Page 89 - Surcharge Mount

Monitor Installation 7-3 Bolting the mounting bracket to the flange welded to the monitor 6. Carefully lower the monitor into the manhole, and plac

Page 90 - Surcharge mount

1-12 ADS FlowShark IS Manual Intrinsically-safe equipment label European ATEX Hazardous Area Compliance The following instructions apply to equipme

Page 91

7-4 ADS FlowShark IS Manual 8. Neatly coil and secure the excess sensor and communication cables in the manhole to simplify future monitor service a

Page 92 - ¾-Band Mount

Monitor Installation 7-5 Mounting the Monitor to the Manhole Rim Mount the monitor to the manhole rim using the monitor bracket in the following way:

Page 93

7-6 ADS FlowShark IS Manual Drilling the hole in the manhole rim 4. Use a 3/8-inch x 16 threads-per-inch tap to cut threads in the hole. Twist the

Page 94

Monitor Installation 7-7 5. Chase the threading action at intervals to clear the metal debris by backing the tap almost completely out of the hole a

Page 95 - ½-Band Mount

7-8 ADS FlowShark IS Manual Binding the nuts together with the washer in between 8. Install the stud into the tapped hole using the outer nut to en

Page 96

Monitor Installation 7-9 Positioning the nuts on the stud to receive the mounting bracket 11. Mount the bottom of the mounting bracket onto the mon

Page 97

8-1 C H A P T E R 8 Maintenance and Troubleshooting While the ADS® FlowShark® IS flow monitor and sensors are designed for dependability and dura

Page 98

8-2 ADS FlowShark IS Manual Maintaining the System Components The FlowShark IS flow monitor and sensors should receive routine on-site inspections a

Page 99

Maintenance and Troubleshooting 8-3 Inspecting the Monitor Perform the following inspections during site visits or from a remote location (when

Page 100 - Pipe and Manhole

8-4 ADS FlowShark IS Manual Checking the Monitor Battery Voltage Check the battery voltage using Profile before installing the monitor and after col

Page 101

Introduction 1-13  If the equipment is likely to come into contact with aggressive substances, the user must take suitable precautions to prevent

Page 102 - Monitor

Maintenance and Troubleshooting 8-5 Monitor Status report indicating Internal and External Voltage Replacing the Monitor Battery Pack Note: Th

Page 103

8-6 ADS FlowShark IS Manual Warning: The FlowShark IS Model 5000-IS1 may only use the 9-volt (ADS p/n 106152) or 12-volt (ADS p/n 8000-0043-01/02)

Page 104 - Communication

Maintenance and Troubleshooting 8-7 1. Contact a data analyst to collect the data remotely or collect the data on-site. This will reduce the ri

Page 105 - Hardware

8-8 ADS FlowShark IS Manual FlowShark IS 9-volt battery pack inside canister 7. (applies only to the 9-volt battery pack) Mark the edge of the ca

Page 106

Maintenance and Troubleshooting 8-9 12-volt battery pack with retainer plug  Carefully slide the thick red band securing the PCB to the side

Page 107 - ADS FlowShark IS Manual

8-10 ADS FlowShark IS Manual 10. (applies only to the 12-volt battery pack) Connect the PCB to the replacement battery pack and then secure the PC

Page 108 - Wire ends stripped ¼-inch

Maintenance and Troubleshooting 8-11 Refer to Replacing EMU Components on page 8-14 for instructions on replacing the battery pack in the EMU (f

Page 109

8-12 ADS FlowShark IS Manual Replacing the Dryer Tube Replace the pressure depth sensor dryer tube in the following way: 1. Clip the cable ties

Page 110

Maintenance and Troubleshooting 8-13 7. Evaluate the pressure depth sensor performance for a limited time to ensure no damage has occurred to t

Page 111 - Installing the Wireless EMU

8-14 ADS FlowShark IS Manual Note: Some desiccant may be reused following a special process designed to remove the moisture from the beads. Please

Page 112

ii ADS FlowShark IS Manual  2011 ADS LLC. All rights reserved. ADS, FlowShark, Profile, ADS Model 3600, and MLI are registered trademarks o

Page 113

1-14 ADS FlowShark IS Manual FlowShark IS ATEX Certificate - Page 1

Page 114

Maintenance and Troubleshooting 8-15 5. Once maintenance is complete, re-mount and secure the EMU to the panel inside the enclosure using the f

Page 115

8-16 ADS FlowShark IS Manual 5. Carefully disconnect the black (power) and the ribbon (communication) cables from the back of the modem. 6. Slide

Page 116 - Interface/Barrier Box

Maintenance and Troubleshooting 8-17 6. Slide the lock to the left to secure the SIM card in the socket. 7. Restore power to the EMU. 8. Cont

Page 117

8-18 ADS FlowShark IS Manual Troubleshooting The FlowShark IS flow monitor contains several different components that perform many different functio

Page 118 - Installing the EMU

Maintenance and Troubleshooting 8-19 Some problems that occur will not require a site visit, such as incorrect equipment identification numbers

Page 119

8-20 ADS FlowShark IS Manual General Monitor Problems The following tables contain general troubleshooting techniques for the FlowShark IS flow moni

Page 120

Maintenance and Troubleshooting 8-21 Problem Modem connects but monitor does not respond. Possible Causes Reset board failure may have occurred

Page 121 - Serial Cable

8-22 ADS FlowShark IS Manual Problem Time on the monitor clock is incorrect. Possible Causes Monitor clock may be a faulty. PC clock may read incorr

Page 122 - Reset PCB)

Maintenance and Troubleshooting 8-23 Problem Data is missing at the beginning or end of the date range following data collection. Possible Cause

Page 123

8-24 ADS FlowShark IS Manual Problem No DC power is available at the EMU. Possible Cause Breaker may have tripped. Station power may be off. Power s

Page 124 - External Power

Introduction 1-15 FlowShark IS ATEX Certificate - Page 2

Page 125 - Typical Installation

Maintenance and Troubleshooting 8-25 Problem Range from ultrasonic depth subsystem is significantly greater than manually measured range. Possib

Page 126 - ADS 12VDC power supply

8-26 ADS FlowShark IS Manual Problem Raw ultrasonic depth data shows depths greater than the pipe diameter. Possible Causes Physical offset may be i

Page 127

Maintenance and Troubleshooting 8-27 Problem Ultrasonic depth readings are erratic. Possible Causes Noise may exist in the sewer pipe. Sensor ma

Page 128

8-28 ADS FlowShark IS Manual Problem Monitor often provides a velocity sensor reading of 0. Possible Causes Sensor may be dirty or broken. Cabling b

Page 129 - Receive External Power

Maintenance and Troubleshooting 8-29 Problem Velocity sensor data seems erratic. Possible Causes Sensor may be dirty. Sensor may be faulty. Anal

Page 130 - Configuration and Activation

8-30 ADS FlowShark IS Manual Problem Temperature reading from the pressure depth sensor shows a value of -273° C (-459° F). Possible Causes Cable co

Page 131

Maintenance and Troubleshooting 8-31 Problem Pressure depth sensor temperature readings are incorrect, but depth data is accurate. Possible Caus

Page 132 - Profile main screen

A-1 A P P E N D I X A Specifications This appendix contains specifications for the ADS FlowShark IS flow monitor and associated subassemblies, sen

Page 133 - Creating a Monitor Location

A-2 ADS FlowShark IS Manual Internal Power One ADS 9-volt or 12-volt IS alkaline battery pack powering the monitor and one ADS non-IS 12-volt alkalin

Page 134

Specifications A-3 IS Digital Signal Processor Board (5000-0636) Processor Texas Instruments VC33 DSP (Digital Signal Processor) Memory 1 megabyte non

Page 135

1-16 ADS FlowShark IS Manual FlowShark IS ATEX Certificate - Page 3

Page 136

A-4 ADS FlowShark IS Manual IS Analog Signal Processor Board (5000-0637) Inputs/Outputs • ADS IS ultrasonic sensor input • ADS IS velocity sensor i

Page 137 - 6-8 ADS FlowShark IS Manual

Specifications A-5 Voltage to Frequency Board (5000-0661) Protection Agency-approved optoisolators; current limiting resistors Construction UL-Recogn

Page 138

A-6 ADS FlowShark IS Manual Antenna Hi-Gain Center-fed Capacitive Monopole Protocol TCP/IP Addressing Fixed address through AT&T Custom APN Netwo

Page 139

Specifications A-7 IS PC Communication Interface Cable (106227A–xxx, where xxx represents length) Size Jacket - 0.318 inches (8 mm) in outside diamet

Page 140

A-8 ADS FlowShark IS Manual IS Pressure Depth Sensor (3754-0001-40) Enclosure Solid machined PVC/stainless steel Dimensions 5.25 inches (13.34 cm) lo

Page 141 - Edit Devices dialog

Specifications A-9 IS Doppler Velocity Sensor (106194A-35) Enclosure Solid machined PVC/epoxy Dimensions 2.35 inches (5.97 cm) long x 1.13 inches (2.

Page 142

B-1 A P P E N D I X B Part Numbers This appendix contains a list of the part numbers for the most commonly used and ordered ADS® FlowShark® IS pa

Page 143

B-2 ADS FlowShark IS Manual 3754-0002-40 SENSOR, PRESSURE, IS, 15PSI, 40' 3754-0003-40 SENSOR, PRESSURE, IS, 30PSI, 40' 106228A CABLE,

Page 144 - Editing the Velocity Device

Part Numbers B-3 EMU Replacement Battery Pack 5000-0557 BATTERY PACK, 24-CELL, ADS5000 EMU Replacement Modem 507209 MODEM, GPRS, QUAD-BAND N

Page 145

In-1 Index 1 1/2-band mount, 3-38 3 3/4-band mount, 3-35 A activation logs, 6-45 ADS Client Services Department, 1-7 ADS Power Supply installation, 5

Page 146

Introduction 1-17 FlowShark IS ATEX Certificate – Page 4

Page 147 - Editing the Pressure Device

In-2 ADS FlowShark IS Manual communication, 4-1 direct, 2-2 setting parameters, 6-20 viewing logs, 6-45 wireless, 2-2, 2-5, 4-2 communication cable,

Page 148 - Editing the MLI Device

Index In-3 EMU (External Modem Unit), 1-4, 4-18 antenna, 4-8 battery pack, 4-8 battery pack replacement, 8-15 battery-powered, 5-6 components, 4-8

Page 149 - Parameters

In-4 ADS FlowShark IS Manual Intrinsic Safety (IS), 1-2 certified tools, 1-5 maintenance, 1-7 IS analog processor board specifications, A-4 IS batte

Page 150

Index In-5 collecting data, 6-34 configuration, 6-2 connecting sensors, 3-45 connectors, 2-10 description, 2-4 digital signal processor board, 2-5 G

Page 151

In-6 ADS FlowShark IS Manual Q QPS Evaluation Services, 1-11 quad-band TCP/IP wireless modem, 4-8 quadredundancy, 2-13 R R&TTE-CE Information, 1

Page 152

Index In-7 special installation, 3-4 Doppler velocity sensor, 3-35 pressure depth sensor, 3-35 sensors, 3-25 ultrasonic depth sensor standard mount,

Page 153 - Activating the Monitor

In-8 ADS FlowShark IS Manual W warranty, 1-35 invalidating, 1-8 wet well, 3-4 wireless communication, 2-5, 4-2 connection, 1-1 EMU, 1-4, 4-8 EMU mod

Page 154 - Monitor Diagnostics dialog

1-18 ADS FlowShark IS Manual FlowShark IS ATEX Certificate – Annex Page 1

Page 155 - Running Device Diagnostics

Introduction 1-19 FlowShark IS ATEX Certificate – Annex Page 2

Page 156

1-20 ADS FlowShark IS Manual 106228 PC Comm Cable ATEX Certificate - Page 1

Page 157

Introduction 1-21 106228 PC Comm Cable ATEX Certificate - Page 2

Page 158

1-22 ADS FlowShark IS Manual 106124 Comm Interface ATEX Certificate - Page 1

Page 159

Introduction 1-23 106124 Comm Interface ATEX Certificate - Page 2

Page 160

Table of Contents iii Table of Contents Chapter 1 Introduction…….. ... 1-1 Intrinsic Safety ...

Page 161

1-24 ADS FlowShark IS Manual 106124 Comm Interface ATEX Certificate - Page 3

Page 162

Introduction 1-25 106186 Ultrasonic Depth Sensor ATEX Certificate - Page 1

Page 163 - Diagnostics toolbar button

1-26 ADS FlowShark IS Manual 106186 Ultrasonic Depth Sensor ATEX Certificate - Page 2

Page 164

Introduction 1-27 3754-0001 Pressure Sensor ATEX Certificate - Page 1

Page 165

1-28 ADS FlowShark IS Manual 3754-0001 Pressure Sensor ATEX Certificate - Page 2

Page 166

Introduction 1-29 3754-0001 Pressure Sensor ATEX Certificate - Page 3

Page 167

1-30 ADS FlowShark IS Manual 106194 Velocity Sensor ATEX Certificate - Page 1

Page 168

Introduction 1-31 106194 Velocity Sensor ATEX Certificate - Page 2

Page 169

1-32 ADS FlowShark IS Manual Additional Notice of Canadian Emissions Requirements This digital apparatus does not exceed the Class A limits for radi

Page 170

Introduction 1-33 Sample Declaration of Conformity

Page 171

iv ADS FlowShark IS Manual Sensors ... 2-12 IS Ultrasonic Depth Sensor ...

Page 172

1-34 ADS FlowShark IS Manual Installation and Configuration Following is the general procedure for installing and configuring a FlowShark monitor. R

Page 173

Introduction 1-35 Product Warranty This section includes the warranty information for the ADS FlowShark IS flow monitor. New Product Warranty All

Page 174

1-36 ADS FlowShark IS Manual Out-of-Warranty Product Repairs After the new product warranty expires, a product may be returned, at the owner’s prepa

Page 175 - View Logs dialog

2-1 C H A P T E R 2 System Overview The ADS® FlowShark® IS flow monitor uses three flow measurement devices to gather raw flow data: a quadredund

Page 176 - Print button

2-2 ADS FlowShark IS Manual Typical FlowShark IS flow monitoring system installation Communication between the monitor and the user’s office or fie

Page 177 - Monitor Installation

System Overview 2-3 Profile also enables the user to process the flow data, generate graphical and tabular reports, organize data in the user'

Page 178

2-4 ADS FlowShark IS Manual ADS FlowShark IS Flow Monitor The FlowShark IS flow monitor is a waterproof, airtight, cylindrical, marine-grade aluminu

Page 179

System Overview 2-5 Communications ADS offers remote wireless communication for the FlowShark IS monitor. The wireless communication is performed o

Page 180

2-6 ADS FlowShark IS Manual The board allocates portions of memory to firmware (embedded code in the monitor that can be updated remotely) and data

Page 181

System Overview 2-7 Analog Board The Analog board (ADS p/n 5000-0637) controls the sampling of the FlowShark IS sensors. It receives signals from

Page 182 - 7-6 ADS FlowShark IS Manual

Table of Contents v Chapter 6 Configuration and Activation .. 6-1 Configuring the Monitor Location ... 6-2 Launching

Page 183

2-8 ADS FlowShark IS Manual Voltage to Frequency Board The Voltage to Frequency board (ADS p/n 5000-0661) allows the FlowShark IS monitor to safely

Page 184

System Overview 2-9 Voltage Regulator Board The Voltage Regulator board (ADS p/n 5000-0638) regulates and distributes power to the analog and DSP b

Page 185

2-10 ADS FlowShark IS Manual Connectors Connectors located on top of the monitor receive the following cabling and components:  IS ultrasonic dept

Page 186 - Troubleshooting

System Overview 2-11 FlowShark IS 9-volt (left) and 12-volt (right) internal battery packs External Power The monitor (and EMU) also can receiv

Page 187

2-12 ADS FlowShark IS Manual Sensors The FlowShark IS flow monitor uses various sensors to gather raw flow data. The IS ultrasonic and pressure dept

Page 188 - Inspecting the Monitor

System Overview 2-13 Ultrasonic depth sensor sending signals to flow surface to determine range Quadredundancy Each ultrasonic depth sensor contai

Page 189

2-14 ADS FlowShark IS Manual Data Scrubbing Flow conditions and internal structures introduce many potential obstacles to obtaining accurate flow da

Page 190

System Overview 2-15 compensates for temperature using a temperature sensor housed within the pressure depth sensor. IS Doppler Velocity Sensor The

Page 191 - 8000-0043-01) battery packs

3-1 C H A P T E R 3 Sensor Installation and Connection The ADS® FlowShark® IS flow monitor and sensors are primarily designed for monitoring flow

Page 192

3-2 ADS FlowShark IS Manual  Secure and Connect the Cables This procedure involves securing the sensor cables from the ring or band to the monito

Page 193

vi ADS FlowShark IS Manual Voltage to Frequency Board ... A-5 Wireless External Modem Unit ...

Page 194

Sensor Installation and Connection 3-3 Investigating Site Characteristics Before beginning installation activities, conduct a thorough investigatio

Page 195

3-4 ADS FlowShark IS Manual  Higher Elevation than High Water Level that Occurs in Wet Well at Downstream Pump Station Pipes that are below the h

Page 196 - Checking the Sensors

Sensor Installation and Connection 3-5  Manhole Depth Measure the distance from the top of the manhole to the bottom of the manhole invert. For

Page 197 - Replacing the Dryer Tube

3-6 ADS FlowShark IS Manual Installing the Sensors in the Pipe This section includes the procedures for performing sensor installations using one of

Page 198 - Replacing the Desiccant

Sensor Installation and Connection 3-7 Quantity Description ADS Part Number 1 IS Doppler Velocity Sensor with two 4-40 x 5/16-inch stainless ste

Page 199 - Replacing EMU Components

3-8 ADS FlowShark IS Manual 4 The stainless steel ring includes the ring stabilizer bracket (p/n I25-0002), spreader mechanism (p/n I10-0003), and

Page 200 - Replacing the EMU Modem

Sensor Installation and Connection 3-9 assembly process. To assemble a non-overlapping ring, proceed directly to step 4. 1. Insert the spreader m

Page 201

3-10 ADS FlowShark IS Manual Sliding the ring stabilizer onto the ring 3. Slide the ring stabilizer all the way around the ring until it is about

Page 202

Sensor Installation and Connection 3-11 facing the outside of the ring. The side with the backstop should face the inside of the ring. Sliding the

Page 203

3-12 ADS FlowShark IS Manual Moving the ring stabilizer into position 9. Perform the following based on the ring type:  Overlapping Insert the

Page 204

1-1 C H A P T E R 1 Introduction The ADS® Intrinsically Safe FlowShark® long-term flow monitor (FlowShark IS) measures open channel flow in sanitar

Page 205 - General Monitor Problems

Sensor Installation and Connection 3-13 11. Orient the ring with the downstream edge (edge with small holes) facing you. 12. Lay the spreader mec

Page 206

3-14 ADS FlowShark IS Manual Attaching the spreader bars 16. Turn the ring until the spreader mechanism is in the 12:00 position. 17. Align the s

Page 207

Sensor Installation and Connection 3-15 Tightening the spreader mechanism screw and hex nut 19. Insert the second spreader mechanism screw throug

Page 208

3-16 ADS FlowShark IS Manual Place the washers onto the second spreader mechanism screw 21. Place the spreader bars over the screw, and tighten a

Page 209

Sensor Installation and Connection 3-17 Although the spreader mechanism fits snugly against the inside of the ring, the spreader mechanism may seem

Page 210

3-18 ADS FlowShark IS Manual  Pressure Depth Sensor Bottom of the ring slightly to the left of the Doppler velocity sensor (when present) Proper

Page 211

Sensor Installation and Connection 3-19 Mounting the ultrasonic depth sensor to the mounting plate on the ring 2. Verify that the ultrasonic dept

Page 212

3-20 ADS FlowShark IS Manual 2. Secure the sensor cable to the ring. Refer to Securing the Cables to the Ring on page 3-21 for instructions on prop

Page 213

Sensor Installation and Connection 3-21 Securing the Cables to the Ring Securing the Doppler velocity sensor and pressure depth sensor cables to th

Page 214 - IS Pressure Depth Subsystem

3-22 ADS FlowShark IS Manual addition, make sure the connector-end of the sensor is not kinked, does not contain moisture, and includes an attached

Page 215

1-2 ADS FlowShark IS Manual Intrinsic Safety Intrinsic Safety is a protection classification for equipment that ensures there are no conditions unde

Page 216

Sensor Installation and Connection 3-23 Keep the following in mind:  Make sure the ultrasonic depth sensor is at the top (crown) of the pipe, the

Page 217 - Specifications

3-24 ADS FlowShark IS Manual Leveling the ultrasonic depth sensor If necessary, adjust the level in the following way:  Remove the ultrasonic dep

Page 218

Sensor Installation and Connection 3-25 Warning: Avoid over-tightening the ring. This could bend the crank assembly. 7. Restore any silt moved to

Page 219

3-26 ADS FlowShark IS Manual Performing a special installation involves the following process:  Gathering the parts and supplies  Gathering the

Page 220

Sensor Installation and Connection 3-27 Quantity Description ADS Part Number 1 IS Doppler velocity sensor with two 4-40 x 5/16-inch stainless st

Page 221 - 3800-0148 EP)

3-28 ADS FlowShark IS Manual 1 Additional models of the FlowShark IS monitor are also available: 5000-IS1-US (without internal battery), 5000-IS1 (f

Page 222 - ADS Power Supply (5000-0727)

Sensor Installation and Connection 3-29 Mounting the Ultrasonic Depth Sensor Two special installation methods are available for mounting the ultras

Page 223 - Intrinsically-Safe Sensors

3-30 ADS FlowShark IS Manual Securing one end of the ultrasonic depth sensor mounting band 6. Slide the adjustable ultrasonic plate onto the band

Page 224

Sensor Installation and Connection 3-31 Securing the other end of the band Note: If the plate has too much play, attach another anchor bolt appro

Page 225

3-32 ADS FlowShark IS Manual 13. Measure the physical offset for the ultrasonic depth sensor (the distance from the crown (top) of the pipe to the

Page 226 - Part Numbers

Introduction 1-3 FlowShark IS System Description The FlowShark IS unit is approved for use only with approved ADS IS sensors, communication cables

Page 227 - B-2 ADS FlowShark IS Manual

Sensor Installation and Connection 3-33 Surcharge mount 2. Press the bracket firmly against the manhole wall. 3. Identify the three pre-drilled

Page 228 - Part Numbers B-3

3-34 ADS FlowShark IS Manual 10. Secure the bottom of the bracket with a washer and nut. 11. Slide the sensor into the bracket, and secure the sen

Page 229

Sensor Installation and Connection 3-35 Therefore, the ¾-band mount is the preferable method for mounting the sensors. However, a ½-band mount may

Page 230

3-36 ADS FlowShark IS Manual 2. Determine which end of the band will be the long end (end extending almost completely up one side of the pipe). For

Page 231 - Index In-3

Sensor Installation and Connection 3-37 7. Place the band in the pipe upstream at least 12 inches (same distance as ultrasonic depth sensor, when

Page 232

3-38 ADS FlowShark IS Manual Sensors mounted using a ¾-band mount in the pipe with and without silt 15. Measure the physical offset for the pressu

Page 233 - Index In-5

Sensor Installation and Connection 3-39 3. Mount the sensor onto the inside of the band with two 4-40 x 5/16-inch countersink screws, making sure

Page 234

3-40 ADS FlowShark IS Manual 9. Install a ¼-inch x 2 ¼-inch anchor bolt, washer, and nut below the flow surface 1 to 3 inches away from the Doppler

Page 235 - Index In-7

Sensor Installation and Connection 3-41 Securing the Cables to the Band Securing the Doppler velocity sensor and pressure depth sensor cables to th

Page 236

3-42 ADS FlowShark IS Manual the ties or kinking the cable can damage or restrict the air tube, causing incorrect pressure depth readings. In additi