ADS Intrinsically-Safe FlowShark Installation, Operation, and Maintenance Manual March 2011 QR 775003 A2 4940 Res
1-4 ADS FlowShark IS Manual Note: A legacy Rechargeable Battery Pack (RBP) may be connected to the COMM connector on the FlowShark IS in a similar
Sensor Installation and Connection 3-43 Securing the Sensor Cables in the Pipe and Manhole Securing the sensor cables from the ring (or band, when
3-44 ADS FlowShark IS Manual 2. Secure the cables with a cable tie and plastic anchor every 18 to 24 inches along the pipe crown (when necessary) f
Sensor Installation and Connection 3-45 Connecting the Sensors to the Monitor After installing the ring and sensors and securing the sensor cables
3-46 ADS FlowShark IS Manual 5. Connect each sensor cable to the appropriate port. Tighten each connector in a clockwise direction until it clicks
4-1 C H A P T E R 4 Communication After installing the sensors in the pipe and connecting the sensors to the ADS® FlowShark® IS monitor, it is nece
4-2 ADS FlowShark IS Manual Installing the Monitor Communication Hardware Establishing remote, wireless communications with the FlowShark IS monitor
Communication 4-3 Verify that the communication cable trench will not damage an existing sprinkler system. Therefore, locate any sprinkler lines.
4-4 ADS FlowShark IS Manual Gathering Tools and Materials Obtain the following items to perform the installation: Ratchet set with 7/16-inch socke
Communication 4-5 Wire ends stripped ¼-inch Cut off the old stripped ends (when applicable) back to the jacket. Use the small side cutters to
4-6 ADS FlowShark IS Manual Forming a thin taper using vinyl tape 4. Excavate a trench at least 12 inches (31 cm) deep and 4 inches (10 cm) wide fr
Introduction 1-5 Only limited field repairs to the FlowShark IS are allowed; all other repairs must be performed by an ADS authorized technician
Communication 4-7 Use urethane foam to seal the space between the conduit and the corbel or manhole wall to prevent infiltration into the manhole.
4-8 ADS FlowShark IS Manual Installing the Wireless EMU Installing the wireless EMU primarily involves wiring the communication cable to the Communic
Communication 4-9 Gathering Tools and Materials Obtain the following items to perform the installation: Medium (No. 2) Phillips head screwdriver
4-10 ADS FlowShark IS Manual Inside component view of EMU box 1. Unscrew the 4 large plastic screws on the cover of the EMU box by hand and remove
Communication 4-11 3. Loosen the nut on the communication cable entry fitting. If the fitting is snapped closed, use a small flat-blade screwdriver
4-12 ADS FlowShark IS Manual 5. Terminate the 12 wires into the orange connector in the following way using the ⅛-inch flat screwdriver. This inform
Communication 4-13 Connecting the communication cable to port on the Communication Interface/Barrier Box 7. Tighten the nut on the communication ca
4-14 ADS FlowShark IS Manual SIM card seated in EMU modem 10. (applies only to battery-powered EMUs) Re-install the battery pack in the EMU in th
Communication 4-15 Installing the EMU Installing the EMU involves mounting the EMU inside the customer-supplied enclosure, connecting the communicati
4-16 ADS FlowShark IS Manual 4. (applies only when using external power) Run and connect the external power cable to the power supply, and then con
1-6 ADS FlowShark IS Manual • Only ADS approved sensors (ADS p/n 106186, 106194, and 3754-0001/0002/0003) shall be connected to the ADS Model 5000-
Communication 4-17 Connecting to the Monitor in the Field Communicating with the monitor on-site occurs by connecting directly to the monitor from a
4-18 ADS FlowShark IS Manual Direct connection using the IS PC communication cable 3. Use the Profile software to establish communication with the
Communication 4-19 EMU serial cable (ADS p/n 3800-0147) Communicate with the monitor through the EMU in the following way: 1. Disconnect the ribbo
4-20 ADS FlowShark IS Manual 3. Connect the other (metal) end of the cable to the serial port on the laptop computer. Direct connection using EMU s
5-1 C H A P T E R 5 External Power The ADS® FlowShark® IS system (monitor and EMU) can receive power from an external DC power source connected to
5-2 ADS FlowShark IS Manual Typical Installation A typical external power installation includes the following: A single-phase, 120-volt AC power l
External Power 5-3 Installing an ADS Power Supply The ADS 12VDC Power Supply (p/n 5000-0727) is a universal input supply that automatically adjusts t
5-4 ADS FlowShark IS Manual Note: Do not drill new holes through the housing for mounting! This will expose the internal components of the enclosure
External Power 5-5 Wiring the external power cable to the ADS Power Supply 6. Verify the following before connecting to the power source: Batter
5-6 ADS FlowShark IS Manual 9. Securely tighten the cable grip fittings and replace the enclosure cover. Make sure the cover properly aligns with th
Introduction 1-7 Maintenance Restrictions As mentioned earlier, all ADS FlowShark IS flow monitors are manufactured to meet IS standards. ADS can v
6-1 C H A P T E R 6 Configuration and Activation After installing the ADS® FlowShark® IS flow monitor and sensors and setting up communication, i
6-2 ADS FlowShark IS Manual Configuring the Monitor Location To ensure the most efficient and useful results for each monitor location, the user mus
Configuration and Activation 6-3 Once installed, start the Profile software from an office or field computer by selecting Start > Programs &g
6-4 ADS FlowShark IS Manual Creating a Monitor Location Creating a new location includes entering and setting up the location information in the fol
Configuration and Activation 6-5 3. Enter the new Location Name. Monitor names must follow this convention depending on length: Use unique n
6-6 ADS FlowShark IS Manual Note: The start and end times displayed in the Monitor Data section represent the complete date/time range of flow data
Configuration and Activation 6-7 1. Select the monitoring point for which you want to create the installation. Monitoring Point 1 selected fro
6-8 ADS FlowShark IS Manual Create Installation – Installation dialog 4. Select the Pipe radio button, and then click on the Next button. Create
Configuration and Activation 6-9 Create Installation – Dimension/Parameters dialog 6. Enter the proper pipe dimensions in the corresponding fi
6-10 ADS FlowShark IS Manual 7. Enter an appropriate name for the installation in the text field, and then click on the Next button. Create Instal
1-8 ADS FlowShark IS Manual Warnings, Certifications, FCC Compliance, and Conformity Manhole and sewer system work involves confined space entry and
Configuration and Activation 6-11 Installation Generator dialog 10. Select File > Exit to close the Installation Generator. Selecting and E
6-12 ADS FlowShark IS Manual Devices selected for HSV4 3. Select the Edit > Properties option or click on the Properties toolbar button. Prope
Configuration and Activation 6-13 device in the Monitoring Point Devices section to ensure Profile includes the device in the LIF. Deselect the
6-14 ADS FlowShark IS Manual Editing the Ultrasonic Device Edit the ultrasonic device in the following way, and then click on the OK button when com
Configuration and Activation 6-15 from the sensor face to the bottom of the pipe. This (default) setting effectively represents the vast majori
6-16 ADS FlowShark IS Manual Note: Typically, the velocity device should not require editing. However, when editing is necessary, ADS recommends ac
Configuration and Activation 6-17 Max Velocity Select the appropriate option from this drop-down list to fine tune the maximum velocity valu
6-18 ADS FlowShark IS Manual Editing the Pressure Device Edit the pressure device in the following way, and then click on the OK button when complet
Configuration and Activation 6-19 Access the ADS website at www.adsenv.com on the Internet. Select ADS Environmental Services > Support
6-20 ADS FlowShark IS Manual Setting the Communication Parameters Setting the communication parameters involves designating the communication ports,
Introduction 1-9 Control Drawing This drawing depicts the interconnections allowed for the FlowShark IS. It is intended for use by inspection profe
Configuration and Activation 6-21 2. Select Communication Parameters from the Functions drop-down list, and click on the Perform button. The Se
6-22 ADS FlowShark IS Manual Attempts Enter the number of times you want your local PC to request data from the monitor following a failed attem
Configuration and Activation 6-23 an external power source, below which you want Profile to provide notification for the FlowShark IS monitor. A
6-24 ADS FlowShark IS Manual Activating the Monitor After configuring the monitor, activate the monitor to initiate the flow monitoring process base
Configuration and Activation 6-25 Monitor Diagnostics dialog 2. Click on the Connect button to establish communication with the monitor. Profi
6-26 ADS FlowShark IS Manual Running Device Diagnostics Profile’s diagnostics tool enables the user to verify the proper operation of sensors, obtai
Configuration and Activation 6-27 2. Click on the Connect button to establish communication with the monitor. Profile initiates communication w
6-28 ADS FlowShark IS Manual Diagnostics dialog showing results from ultrasonic depth sensor firing Click on the Advanced button to view more d
Configuration and Activation 6-29 Click on the Close button to exit the Ultrasonic 1 device dialog and return to the Monitor Diagnostics dial
6-30 ADS FlowShark IS Manual Diagnostics dialog showing results from velocity sensor firing Click on the Advanced button to view more detailed d
1-10 ADS FlowShark IS Manual These limits are designed to provide reasonable protection against harmful interference in a residential installation.
Configuration and Activation 6-31 Click on the Close button to exit the Velocity 1 device dialog and return to the Monitor Diagnostics dialog
6-32 ADS FlowShark IS Manual Diagnostics dialog showing results from pressure depth sensor firing Click on the Advanced button to view more deta
Configuration and Activation 6-33 7. Click on the Disconnect button to discontinue communication with the monitor when finished running diagnos
6-34 ADS FlowShark IS Manual Collecting Data from the Monitor The data collection function available through Diagnostics provides enhanced features
Configuration and Activation 6-35 Monitor Diagnostics dialog 2. Click on the Connect button to establish communication with the monitor. Profi
6-36 ADS FlowShark IS Manual Collect the Specified Data dialog 4. Designate the range of data you want to collect from the monitor by editing the
Configuration and Activation 6-37 Collect The Specified Data dialog indicating data collection was a success 6. Click on the Next button. The
6-38 ADS FlowShark IS Manual View Hydrograph – [location name] dialog 7. Review the data on the hydrograph, and then click on the Next button. Und
Configuration and Activation 6-39 View Depth and Velocity Scattergraph – [location name] dialog 8. Review the data on the scattergraph, and th
6-40 ADS FlowShark IS Manual View Ultrasonic and Pressure Scattergraph – [location name] dialog 9. Review the data on the scattergraph, and then c
Introduction 1-11 R&TTE-CE Information U.S. Hazardous Area Compliance The FlowShark IS has been certified for use in Class I, Division 1, Gr
Configuration and Activation 6-41 Site DR Analysis – [location name] dialog 10. Review any issues identified and the suggested actions, and th
6-42 ADS FlowShark IS Manual Upgrading the Monitor Firmware Profile enables you to download updated firmware to the FlowShark IS flow monitor that m
Configuration and Activation 6-43 Monitor Diagnostics – [location name] dialog 2. Click on the Connect button to establish communication with
6-44 ADS FlowShark IS Manual Form Firmware Download dialog 4. Click on the Browse button corresponding to DSP file to locate and designate the DSP
Configuration and Activation 6-45 Viewing Diagnostic and Data Logs Profile generates detailed logs for many activities performed through Diagnos
6-46 ADS FlowShark IS Manual 2. Select the monitoring point for which you want to view a log from the Monitoring Point drop-down list. 3. Select L
Configuration and Activation 6-47 Example of Monitor Status log 7. (optional) Click on the Print button to print the contents of the log file
7-1 C H A P T E R 7 Monitor Installation After installing the sensors and establishing communications, install the ADS® FlowShark® IS monitor in t
7-2 ADS FlowShark IS Manual Mounting the Monitor to the Manhole Wall Mount the monitor to the manhole wall using the monitor bracket in the following
Monitor Installation 7-3 Bolting the mounting bracket to the flange welded to the monitor 6. Carefully lower the monitor into the manhole, and plac
1-12 ADS FlowShark IS Manual Intrinsically-safe equipment label European ATEX Hazardous Area Compliance The following instructions apply to equipme
7-4 ADS FlowShark IS Manual 8. Neatly coil and secure the excess sensor and communication cables in the manhole to simplify future monitor service a
Monitor Installation 7-5 Mounting the Monitor to the Manhole Rim Mount the monitor to the manhole rim using the monitor bracket in the following way:
7-6 ADS FlowShark IS Manual Drilling the hole in the manhole rim 4. Use a 3/8-inch x 16 threads-per-inch tap to cut threads in the hole. Twist the
Monitor Installation 7-7 5. Chase the threading action at intervals to clear the metal debris by backing the tap almost completely out of the hole a
7-8 ADS FlowShark IS Manual Binding the nuts together with the washer in between 8. Install the stud into the tapped hole using the outer nut to en
Monitor Installation 7-9 Positioning the nuts on the stud to receive the mounting bracket 11. Mount the bottom of the mounting bracket onto the mon
8-1 C H A P T E R 8 Maintenance and Troubleshooting While the ADS® FlowShark® IS flow monitor and sensors are designed for dependability and dura
8-2 ADS FlowShark IS Manual Maintaining the System Components The FlowShark IS flow monitor and sensors should receive routine on-site inspections a
Maintenance and Troubleshooting 8-3 Inspecting the Monitor Perform the following inspections during site visits or from a remote location (when
8-4 ADS FlowShark IS Manual Checking the Monitor Battery Voltage Check the battery voltage using Profile before installing the monitor and after col
Introduction 1-13 If the equipment is likely to come into contact with aggressive substances, the user must take suitable precautions to prevent
Maintenance and Troubleshooting 8-5 Monitor Status report indicating Internal and External Voltage Replacing the Monitor Battery Pack Note: Th
8-6 ADS FlowShark IS Manual Warning: The FlowShark IS Model 5000-IS1 may only use the 9-volt (ADS p/n 106152) or 12-volt (ADS p/n 8000-0043-01/02)
Maintenance and Troubleshooting 8-7 1. Contact a data analyst to collect the data remotely or collect the data on-site. This will reduce the ri
8-8 ADS FlowShark IS Manual FlowShark IS 9-volt battery pack inside canister 7. (applies only to the 9-volt battery pack) Mark the edge of the ca
Maintenance and Troubleshooting 8-9 12-volt battery pack with retainer plug Carefully slide the thick red band securing the PCB to the side
8-10 ADS FlowShark IS Manual 10. (applies only to the 12-volt battery pack) Connect the PCB to the replacement battery pack and then secure the PC
Maintenance and Troubleshooting 8-11 Refer to Replacing EMU Components on page 8-14 for instructions on replacing the battery pack in the EMU (f
8-12 ADS FlowShark IS Manual Replacing the Dryer Tube Replace the pressure depth sensor dryer tube in the following way: 1. Clip the cable ties
Maintenance and Troubleshooting 8-13 7. Evaluate the pressure depth sensor performance for a limited time to ensure no damage has occurred to t
8-14 ADS FlowShark IS Manual Note: Some desiccant may be reused following a special process designed to remove the moisture from the beads. Please
ii ADS FlowShark IS Manual 2011 ADS LLC. All rights reserved. ADS, FlowShark, Profile, ADS Model 3600, and MLI are registered trademarks o
1-14 ADS FlowShark IS Manual FlowShark IS ATEX Certificate - Page 1
Maintenance and Troubleshooting 8-15 5. Once maintenance is complete, re-mount and secure the EMU to the panel inside the enclosure using the f
8-16 ADS FlowShark IS Manual 5. Carefully disconnect the black (power) and the ribbon (communication) cables from the back of the modem. 6. Slide
Maintenance and Troubleshooting 8-17 6. Slide the lock to the left to secure the SIM card in the socket. 7. Restore power to the EMU. 8. Cont
8-18 ADS FlowShark IS Manual Troubleshooting The FlowShark IS flow monitor contains several different components that perform many different functio
Maintenance and Troubleshooting 8-19 Some problems that occur will not require a site visit, such as incorrect equipment identification numbers
8-20 ADS FlowShark IS Manual General Monitor Problems The following tables contain general troubleshooting techniques for the FlowShark IS flow moni
Maintenance and Troubleshooting 8-21 Problem Modem connects but monitor does not respond. Possible Causes Reset board failure may have occurred
8-22 ADS FlowShark IS Manual Problem Time on the monitor clock is incorrect. Possible Causes Monitor clock may be a faulty. PC clock may read incorr
Maintenance and Troubleshooting 8-23 Problem Data is missing at the beginning or end of the date range following data collection. Possible Cause
8-24 ADS FlowShark IS Manual Problem No DC power is available at the EMU. Possible Cause Breaker may have tripped. Station power may be off. Power s
Introduction 1-15 FlowShark IS ATEX Certificate - Page 2
Maintenance and Troubleshooting 8-25 Problem Range from ultrasonic depth subsystem is significantly greater than manually measured range. Possib
8-26 ADS FlowShark IS Manual Problem Raw ultrasonic depth data shows depths greater than the pipe diameter. Possible Causes Physical offset may be i
Maintenance and Troubleshooting 8-27 Problem Ultrasonic depth readings are erratic. Possible Causes Noise may exist in the sewer pipe. Sensor ma
8-28 ADS FlowShark IS Manual Problem Monitor often provides a velocity sensor reading of 0. Possible Causes Sensor may be dirty or broken. Cabling b
Maintenance and Troubleshooting 8-29 Problem Velocity sensor data seems erratic. Possible Causes Sensor may be dirty. Sensor may be faulty. Anal
8-30 ADS FlowShark IS Manual Problem Temperature reading from the pressure depth sensor shows a value of -273° C (-459° F). Possible Causes Cable co
Maintenance and Troubleshooting 8-31 Problem Pressure depth sensor temperature readings are incorrect, but depth data is accurate. Possible Caus
A-1 A P P E N D I X A Specifications This appendix contains specifications for the ADS FlowShark IS flow monitor and associated subassemblies, sen
A-2 ADS FlowShark IS Manual Internal Power One ADS 9-volt or 12-volt IS alkaline battery pack powering the monitor and one ADS non-IS 12-volt alkalin
Specifications A-3 IS Digital Signal Processor Board (5000-0636) Processor Texas Instruments VC33 DSP (Digital Signal Processor) Memory 1 megabyte non
1-16 ADS FlowShark IS Manual FlowShark IS ATEX Certificate - Page 3
A-4 ADS FlowShark IS Manual IS Analog Signal Processor Board (5000-0637) Inputs/Outputs • ADS IS ultrasonic sensor input • ADS IS velocity sensor i
Specifications A-5 Voltage to Frequency Board (5000-0661) Protection Agency-approved optoisolators; current limiting resistors Construction UL-Recogn
A-6 ADS FlowShark IS Manual Antenna Hi-Gain Center-fed Capacitive Monopole Protocol TCP/IP Addressing Fixed address through AT&T Custom APN Netwo
Specifications A-7 IS PC Communication Interface Cable (106227A–xxx, where xxx represents length) Size Jacket - 0.318 inches (8 mm) in outside diamet
A-8 ADS FlowShark IS Manual IS Pressure Depth Sensor (3754-0001-40) Enclosure Solid machined PVC/stainless steel Dimensions 5.25 inches (13.34 cm) lo
Specifications A-9 IS Doppler Velocity Sensor (106194A-35) Enclosure Solid machined PVC/epoxy Dimensions 2.35 inches (5.97 cm) long x 1.13 inches (2.
B-1 A P P E N D I X B Part Numbers This appendix contains a list of the part numbers for the most commonly used and ordered ADS® FlowShark® IS pa
B-2 ADS FlowShark IS Manual 3754-0002-40 SENSOR, PRESSURE, IS, 15PSI, 40' 3754-0003-40 SENSOR, PRESSURE, IS, 30PSI, 40' 106228A CABLE,
Part Numbers B-3 EMU Replacement Battery Pack 5000-0557 BATTERY PACK, 24-CELL, ADS5000 EMU Replacement Modem 507209 MODEM, GPRS, QUAD-BAND N
In-1 Index 1 1/2-band mount, 3-38 3 3/4-band mount, 3-35 A activation logs, 6-45 ADS Client Services Department, 1-7 ADS Power Supply installation, 5
Introduction 1-17 FlowShark IS ATEX Certificate – Page 4
In-2 ADS FlowShark IS Manual communication, 4-1 direct, 2-2 setting parameters, 6-20 viewing logs, 6-45 wireless, 2-2, 2-5, 4-2 communication cable,
Index In-3 EMU (External Modem Unit), 1-4, 4-18 antenna, 4-8 battery pack, 4-8 battery pack replacement, 8-15 battery-powered, 5-6 components, 4-8
In-4 ADS FlowShark IS Manual Intrinsic Safety (IS), 1-2 certified tools, 1-5 maintenance, 1-7 IS analog processor board specifications, A-4 IS batte
Index In-5 collecting data, 6-34 configuration, 6-2 connecting sensors, 3-45 connectors, 2-10 description, 2-4 digital signal processor board, 2-5 G
In-6 ADS FlowShark IS Manual Q QPS Evaluation Services, 1-11 quad-band TCP/IP wireless modem, 4-8 quadredundancy, 2-13 R R&TTE-CE Information, 1
Index In-7 special installation, 3-4 Doppler velocity sensor, 3-35 pressure depth sensor, 3-35 sensors, 3-25 ultrasonic depth sensor standard mount,
In-8 ADS FlowShark IS Manual W warranty, 1-35 invalidating, 1-8 wet well, 3-4 wireless communication, 2-5, 4-2 connection, 1-1 EMU, 1-4, 4-8 EMU mod
1-18 ADS FlowShark IS Manual FlowShark IS ATEX Certificate – Annex Page 1
Introduction 1-19 FlowShark IS ATEX Certificate – Annex Page 2
1-20 ADS FlowShark IS Manual 106228 PC Comm Cable ATEX Certificate - Page 1
Introduction 1-21 106228 PC Comm Cable ATEX Certificate - Page 2
1-22 ADS FlowShark IS Manual 106124 Comm Interface ATEX Certificate - Page 1
Introduction 1-23 106124 Comm Interface ATEX Certificate - Page 2
Table of Contents iii Table of Contents Chapter 1 Introduction…….. ... 1-1 Intrinsic Safety ...
1-24 ADS FlowShark IS Manual 106124 Comm Interface ATEX Certificate - Page 3
Introduction 1-25 106186 Ultrasonic Depth Sensor ATEX Certificate - Page 1
1-26 ADS FlowShark IS Manual 106186 Ultrasonic Depth Sensor ATEX Certificate - Page 2
Introduction 1-27 3754-0001 Pressure Sensor ATEX Certificate - Page 1
1-28 ADS FlowShark IS Manual 3754-0001 Pressure Sensor ATEX Certificate - Page 2
Introduction 1-29 3754-0001 Pressure Sensor ATEX Certificate - Page 3
1-30 ADS FlowShark IS Manual 106194 Velocity Sensor ATEX Certificate - Page 1
Introduction 1-31 106194 Velocity Sensor ATEX Certificate - Page 2
1-32 ADS FlowShark IS Manual Additional Notice of Canadian Emissions Requirements This digital apparatus does not exceed the Class A limits for radi
Introduction 1-33 Sample Declaration of Conformity
iv ADS FlowShark IS Manual Sensors ... 2-12 IS Ultrasonic Depth Sensor ...
1-34 ADS FlowShark IS Manual Installation and Configuration Following is the general procedure for installing and configuring a FlowShark monitor. R
Introduction 1-35 Product Warranty This section includes the warranty information for the ADS FlowShark IS flow monitor. New Product Warranty All
1-36 ADS FlowShark IS Manual Out-of-Warranty Product Repairs After the new product warranty expires, a product may be returned, at the owner’s prepa
2-1 C H A P T E R 2 System Overview The ADS® FlowShark® IS flow monitor uses three flow measurement devices to gather raw flow data: a quadredund
2-2 ADS FlowShark IS Manual Typical FlowShark IS flow monitoring system installation Communication between the monitor and the user’s office or fie
System Overview 2-3 Profile also enables the user to process the flow data, generate graphical and tabular reports, organize data in the user'
2-4 ADS FlowShark IS Manual ADS FlowShark IS Flow Monitor The FlowShark IS flow monitor is a waterproof, airtight, cylindrical, marine-grade aluminu
System Overview 2-5 Communications ADS offers remote wireless communication for the FlowShark IS monitor. The wireless communication is performed o
2-6 ADS FlowShark IS Manual The board allocates portions of memory to firmware (embedded code in the monitor that can be updated remotely) and data
System Overview 2-7 Analog Board The Analog board (ADS p/n 5000-0637) controls the sampling of the FlowShark IS sensors. It receives signals from
Table of Contents v Chapter 6 Configuration and Activation .. 6-1 Configuring the Monitor Location ... 6-2 Launching
2-8 ADS FlowShark IS Manual Voltage to Frequency Board The Voltage to Frequency board (ADS p/n 5000-0661) allows the FlowShark IS monitor to safely
System Overview 2-9 Voltage Regulator Board The Voltage Regulator board (ADS p/n 5000-0638) regulates and distributes power to the analog and DSP b
2-10 ADS FlowShark IS Manual Connectors Connectors located on top of the monitor receive the following cabling and components: IS ultrasonic dept
System Overview 2-11 FlowShark IS 9-volt (left) and 12-volt (right) internal battery packs External Power The monitor (and EMU) also can receiv
2-12 ADS FlowShark IS Manual Sensors The FlowShark IS flow monitor uses various sensors to gather raw flow data. The IS ultrasonic and pressure dept
System Overview 2-13 Ultrasonic depth sensor sending signals to flow surface to determine range Quadredundancy Each ultrasonic depth sensor contai
2-14 ADS FlowShark IS Manual Data Scrubbing Flow conditions and internal structures introduce many potential obstacles to obtaining accurate flow da
System Overview 2-15 compensates for temperature using a temperature sensor housed within the pressure depth sensor. IS Doppler Velocity Sensor The
3-1 C H A P T E R 3 Sensor Installation and Connection The ADS® FlowShark® IS flow monitor and sensors are primarily designed for monitoring flow
3-2 ADS FlowShark IS Manual Secure and Connect the Cables This procedure involves securing the sensor cables from the ring or band to the monito
vi ADS FlowShark IS Manual Voltage to Frequency Board ... A-5 Wireless External Modem Unit ...
Sensor Installation and Connection 3-3 Investigating Site Characteristics Before beginning installation activities, conduct a thorough investigatio
3-4 ADS FlowShark IS Manual Higher Elevation than High Water Level that Occurs in Wet Well at Downstream Pump Station Pipes that are below the h
Sensor Installation and Connection 3-5 Manhole Depth Measure the distance from the top of the manhole to the bottom of the manhole invert. For
3-6 ADS FlowShark IS Manual Installing the Sensors in the Pipe This section includes the procedures for performing sensor installations using one of
Sensor Installation and Connection 3-7 Quantity Description ADS Part Number 1 IS Doppler Velocity Sensor with two 4-40 x 5/16-inch stainless ste
3-8 ADS FlowShark IS Manual 4 The stainless steel ring includes the ring stabilizer bracket (p/n I25-0002), spreader mechanism (p/n I10-0003), and
Sensor Installation and Connection 3-9 assembly process. To assemble a non-overlapping ring, proceed directly to step 4. 1. Insert the spreader m
3-10 ADS FlowShark IS Manual Sliding the ring stabilizer onto the ring 3. Slide the ring stabilizer all the way around the ring until it is about
Sensor Installation and Connection 3-11 facing the outside of the ring. The side with the backstop should face the inside of the ring. Sliding the
3-12 ADS FlowShark IS Manual Moving the ring stabilizer into position 9. Perform the following based on the ring type: Overlapping Insert the
1-1 C H A P T E R 1 Introduction The ADS® Intrinsically Safe FlowShark® long-term flow monitor (FlowShark IS) measures open channel flow in sanitar
Sensor Installation and Connection 3-13 11. Orient the ring with the downstream edge (edge with small holes) facing you. 12. Lay the spreader mec
3-14 ADS FlowShark IS Manual Attaching the spreader bars 16. Turn the ring until the spreader mechanism is in the 12:00 position. 17. Align the s
Sensor Installation and Connection 3-15 Tightening the spreader mechanism screw and hex nut 19. Insert the second spreader mechanism screw throug
3-16 ADS FlowShark IS Manual Place the washers onto the second spreader mechanism screw 21. Place the spreader bars over the screw, and tighten a
Sensor Installation and Connection 3-17 Although the spreader mechanism fits snugly against the inside of the ring, the spreader mechanism may seem
3-18 ADS FlowShark IS Manual Pressure Depth Sensor Bottom of the ring slightly to the left of the Doppler velocity sensor (when present) Proper
Sensor Installation and Connection 3-19 Mounting the ultrasonic depth sensor to the mounting plate on the ring 2. Verify that the ultrasonic dept
3-20 ADS FlowShark IS Manual 2. Secure the sensor cable to the ring. Refer to Securing the Cables to the Ring on page 3-21 for instructions on prop
Sensor Installation and Connection 3-21 Securing the Cables to the Ring Securing the Doppler velocity sensor and pressure depth sensor cables to th
3-22 ADS FlowShark IS Manual addition, make sure the connector-end of the sensor is not kinked, does not contain moisture, and includes an attached
1-2 ADS FlowShark IS Manual Intrinsic Safety Intrinsic Safety is a protection classification for equipment that ensures there are no conditions unde
Sensor Installation and Connection 3-23 Keep the following in mind: Make sure the ultrasonic depth sensor is at the top (crown) of the pipe, the
3-24 ADS FlowShark IS Manual Leveling the ultrasonic depth sensor If necessary, adjust the level in the following way: Remove the ultrasonic dep
Sensor Installation and Connection 3-25 Warning: Avoid over-tightening the ring. This could bend the crank assembly. 7. Restore any silt moved to
3-26 ADS FlowShark IS Manual Performing a special installation involves the following process: Gathering the parts and supplies Gathering the
Sensor Installation and Connection 3-27 Quantity Description ADS Part Number 1 IS Doppler velocity sensor with two 4-40 x 5/16-inch stainless st
3-28 ADS FlowShark IS Manual 1 Additional models of the FlowShark IS monitor are also available: 5000-IS1-US (without internal battery), 5000-IS1 (f
Sensor Installation and Connection 3-29 Mounting the Ultrasonic Depth Sensor Two special installation methods are available for mounting the ultras
3-30 ADS FlowShark IS Manual Securing one end of the ultrasonic depth sensor mounting band 6. Slide the adjustable ultrasonic plate onto the band
Sensor Installation and Connection 3-31 Securing the other end of the band Note: If the plate has too much play, attach another anchor bolt appro
3-32 ADS FlowShark IS Manual 13. Measure the physical offset for the ultrasonic depth sensor (the distance from the crown (top) of the pipe to the
Introduction 1-3 FlowShark IS System Description The FlowShark IS unit is approved for use only with approved ADS IS sensors, communication cables
Sensor Installation and Connection 3-33 Surcharge mount 2. Press the bracket firmly against the manhole wall. 3. Identify the three pre-drilled
3-34 ADS FlowShark IS Manual 10. Secure the bottom of the bracket with a washer and nut. 11. Slide the sensor into the bracket, and secure the sen
Sensor Installation and Connection 3-35 Therefore, the ¾-band mount is the preferable method for mounting the sensors. However, a ½-band mount may
3-36 ADS FlowShark IS Manual 2. Determine which end of the band will be the long end (end extending almost completely up one side of the pipe). For
Sensor Installation and Connection 3-37 7. Place the band in the pipe upstream at least 12 inches (same distance as ultrasonic depth sensor, when
3-38 ADS FlowShark IS Manual Sensors mounted using a ¾-band mount in the pipe with and without silt 15. Measure the physical offset for the pressu
Sensor Installation and Connection 3-39 3. Mount the sensor onto the inside of the band with two 4-40 x 5/16-inch countersink screws, making sure
3-40 ADS FlowShark IS Manual 9. Install a ¼-inch x 2 ¼-inch anchor bolt, washer, and nut below the flow surface 1 to 3 inches away from the Doppler
Sensor Installation and Connection 3-41 Securing the Cables to the Band Securing the Doppler velocity sensor and pressure depth sensor cables to th
3-42 ADS FlowShark IS Manual the ties or kinking the cable can damage or restrict the air tube, causing incorrect pressure depth readings. In additi